2019-09-17 — E1269

Calibration certificate details for 2019-09-17.

Certificate Metadata

Certificate Number
E1269
Calibration Date
2019-09-17
Due Date
Manufacturer
Additel
Model
220
Serial Number
31218020006
Description
Multifunction Loop Calibrator
Customer
Analyt-MTC GmbH
Issuing Lab
Europascal GmbH
Adjusted
No
Remarks
Certificate PDF
Download

DC Voltage Measure (V)

DC Voltage measure, Range 300mV

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
-55-55.00000-54.99920.00098
-0.29-0.29-0.2900070.0000054
0.010.010.010.0000013
0.050.050.0500010.0000018
0.10.10.1000030.0000026
0.290.290.2900080.0000054
11.000001.00000.000059
1515.0000014.99990.00018
2525.0000025.00000.00028
5555.0000054.99970.00098

DC Current Measure (mA)

DC Current measure, Range 30mA

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
44.000003.99990.00061
1212.0000011.99980.0013
2020.0000019.99960.0021
2929.0000028.99930.0029
-29-29.00000-28.99940.0029

DC Voltage Source (V)

Voltage source, Range 200 mV

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
-0.009-0.009001-0.0090.0000058
0.010.0099980.010.0000058
0.0750.0749970.0750.0000058
0.1350.1349980.1350.0000059
0.190.1899980.190.0000061
21.99992.0000.00058
43.99964.0000.00058
65.99966.0000.00058
87.99938.0000.00058
109.999310.0000.00058

DC Current Source (mA)

DC Current source, Range 22mA

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
43.99954.0000.00066
87.99968.0000.00081
1211.999812.0000.0010
1615.999716.0000.0022
2019.999620.0000.0026

DC 24V Output (V)

DC 24V Output

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
2424.0824.00.058