2025-08-20 — E6222

Calibration certificate details for 2025-08-20.

Certificate Metadata

Certificate Number
E6222
Calibration Date
2025-08-20
Due Date
Manufacturer
BK PRECISION
Model
5491A
Serial Number
549111175070801900
Description
Dual Display Multimeter
Customer
Analyt-MTC GmbH
Issuing Lab
Europascal GmbH
Adjusted
No
Remarks
Certificate PDF
Download

DC Voltage (V)

DC Voltage, Range 500mV

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
-490-490.000-490.030.012
-49-49.0000-49.0050.0012
-4.9-4.90000-4.90070.00011
-2.5-2.50000-2.50030.000087
-1-1.00000-1.00010.000059
-0.5-0.50000-0.49990.000058
0000.0000059
00.000000.00010.000058
00.00000.0000.00058
00.0000.000.0058
0.010.010.010.0000059
0.250.250.250020.0000083
0.490.490.490050.0000094
0.50.500000.50010.000058
11.000001.00020.000059
2.52.500002.50050.000065
4.94.900004.90090.000098
4949.000049.0050.0012
490490.000490.040.012
750750.00750.10.060

DC Current (A)

DC Current, Range 500 μA

NominalReference (As Found)DUT (As Found)UncertaintyDev/Spec (As Found)
000.000000020.000000022
0.00040.00040.00040010.00000009
0.0040.0040.00400080.00000066
0.040.040.0400010.0000065
0.40.40.400010.00012
22.000001.99780.00065
22.00001.9980.00087