Molbloc/Molbox
Uncertainty budget group: Molbloc/Molbox.
Uncertainty budgets for Fluke Calibration molbloc/molbox laminar flow reference standards.
Budgets
| Variant | Standard Uncertainty (k=1) | Expanded Uncertainty (k=2) |
|---|---|---|
| Molbloc-L Mass Flow Calibration — A700k, Full Modelization, Low Pressure | 0.107% rdg + 0.0084% FS | 0.214% rdg + 0.0168% FS |